While the customer was using a part of an intraoral sensor system, schick33 s2 sensor ship kit/3.0 cable 6', they allege the device self triggered, causing an x-ray image to be taken without prompting the device to take an image.No injury was reported from the alleged event.
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- problem description: (b)(6) 2026 15:25:44 (b)(4).Initial log (b)(6) 2026 3:22:27 pm (dentsply sirona charlotte time).Warranty: yes.Dealer on-site: no.Billable: no.Billing accepted: billing approved by: 90 day expiration date: 04/13/2026.Remote access: no.Sidexis version: server location: pdata remote location: rcu location (if applicable): practice management: workstation(s): scout check: firmware: plug-in version: exposure count: unit ip address: troubleshooting description: sensor autofires sn (b)(6) not showing in dscrm office has replaced the sensor cable and reseated the elastomer and still having issues suggest replacing sensor under warranty.
|