| Catalog Number |
8065000297 |
| Medical Device Problem Code |
Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that during anterior vitrectomy surgery an ophthalmic operating system had too much fluctuations in the chamber.The surgery was completed on the same day by changing the system for emergency basis.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The company representative was unable to confirm nor replicate the reported issue.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.There were no similar complaints reported for the ¿product serial¿ under investigation, with the same event code.The system was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.A potential root cause of the reported ¿chamber instability issues¿ may be attributed to surgical variables as mentioned above in excerpts from the operator's manual.However, this was not confirmed in the information presented.An internal investigation was opened to address this issue.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Similar event data was collected for the associated reported events.Quality assurance will continue to perform periodic monitoring for evidence of adverse trending and take further action as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|