| Model Number |
BF-P190 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's investigation.The following fields were updated: d9, h2, h3, and h6.The device was returned to olympus for inspection and the reported event was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: damaged mechanism.The most probable cause of the complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reporting during a diagnostic bronchoscopy procedure, the movement lever locked on the bronchovideoscope.There was no patient harm or delay.
|
| |
|