| Catalog Number |
IFNT513 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Osteolysis (2377)
|
| Date of Event |
01/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A4: weight unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost non-existent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was pain and discomfort around the implant.The implant was removed due to peri-implantitis.Symptoms as a result of the event: inflammation & pain.Tooth location was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported event remains non-verifiable as a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection, bone loss, and/or peri-implantitis problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.
|
| |
|
Event or Problem Description
|
|
No additional event information received at the time of this report.
|
| |
|
Search Alerts/Recalls
|