• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 5 FR X 20 CM; CATHETER PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 5 FR X 20 CM; CATHETER PERCUTANEOUS Back to Search Results
Catalog Number ES-04301
Medical Device Problem Codes Break (1069); Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/15/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that " on (b)(6) 2026, upon aspiration for blood sampling via the deep vein catheter in the morning, air was aspirated repeatedly.A breakage was identified at the 18 cm mark of the central venous catheter (cvc).The catheter was then removed and replaced with a new one.The deep vein catheterization was performed on (b)(6)." the user changed to a new set to complete the procedure.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "(b)(6) 2026, upon aspiration for blood sampling via the deep vein catheter in the morning, air was aspirated repeatedly.A breakage was identified at the 18 cm mark of the central venous catheter (cvc).The catheter was then removed and replaced with a new one.The deep vein catheterization was performed on january 13th." the user changed to a new set to complete the procedure.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 5 FR X 20 CM
Common Device Name
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key24314106
Report Number3006425876-2026-00161
Device Sequence Number5676234
Product Code DQY
UDI-Device Identifier10801902184907
UDI-Public(01)10801902184907(11)241205(17)291130(10)71F24L1301(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K862056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberES-04301
Device Lot Number71F24L1301
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2026
Supplement Date Received by Manufacturer03/06/2026
Initial Report FDA Received Date02/11/2026
Supplement Report FDA Received Date03/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED.; NOT REPORTED.
Patient SexUnknown
-
-