Catalog Number ECR60B
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown surgery, could not fire.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent 2/11/2026.D4 batch #148c31.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned reload visual analysis of the returned sample determined that one ecr60b reload was received partially fired 1/4 and with an opened package.The returned reload was reset and loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 148c31, and no non-conformances were identified.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUE RELOAD FOR ECHELON 60
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
3035526892
MDR Report Key24314618
Report Number3005075853-2026-01263
Device Sequence Number17535893
Product Code GDW
UDI-Device Identifier10705036001775
UDI-Public(01)10705036001775(17)270930(10)172C97(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECR60B
Device Lot Number172C97
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2026
Initial Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-