(b)(4).Date sent 2/11/2026.D4 batch #148c31.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned reload visual analysis of the returned sample determined that one ecr60b reload was received partially fired 1/4 and with an opened package.The returned reload was reset and loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 148c31, and no non-conformances were identified.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.
|