• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND ESSENZ; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND ESSENZ; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CONSOLE PLUS
Medical Device Problem Code Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
Livanova deutschland received a report that essenz hlm cockpit froze after case and rebooted.The event occurred at the end of procedure.There is no report of any patient injury.
 
Additional Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H11: livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
H11 complaints database review revealed that similar events have been reported by other customers.Notably, throughout the reboot event , the pumps continued to operate fully, and the system remained controllable through the backup control unit with no impact on the clinical procedure.Reboot is an intended mitigation to reestablish the user interface in case of an unrecoverable exception at the tscp board level.This causes the linux operating system to reset the application to resolve the problem.By design, during the reboot, the heart-lung machine's critical functions, including pumps, alarms, sensors, and safety systems, continue to perform as expected.During the reboot, the cockpit screen shows the livanova logo and once the cockpit user interface is restored, the ¿last case¿ button enables users to retrieve all prior settings and proceed without any data loss.Livanova initiated an improvement action in order to further decrease the already low frequency of occurrence of this type of event.The risk is low and in the acceptable region.Livanova will keep monitoring the market.
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENZ
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
annarita di lascio
14401 w. 65th way
arvada, CO 80004
MDR Report Key24314643
Report Number9611109-2026-90040
Device Sequence Number17555250
Product Code DTQ
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K232291
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCONSOLE PLUS
Device Catalogue Number49-00-10
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2026
Supplement Date Received by Manufacturer03/16/2026
Initial Report FDA Received Date02/11/2026
Supplement Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/30/2025
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-