While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation: solution description on 01/15/2026, 16:08:20, (b)(6).Initial log: on 01/15/2026 4:06:13 pm, (dentsply sirona charlotte time).Warranty: dealer on-site: billable: n, billing accepted: billing approved by: 90 day expiration date: 04/15/2026, remote access: sidexis version: server location: pdata remote location: rcu location (if applicable): practice management: workstation(s): op03, scout check: firmware: plug-in version: exposure count: unit ip address: troubleshooting description: i remoted in unable to retrieve image no images recoverable.Therefore, the image would need to be taken again.
|
|
While the customer was using a part of an intraoral sensor system, schick33 s2 sensor ship kit/3.0 cable 6', they allege there was an x-ray image transmission issue where the x-ray image was not captured after radiation has occurred (transmission failure).No injury was reported from the alleged event.
|