While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- solution description 01/15/2026 ; 17:40:48; (b)(4).--------initial log-------- 01/15/2026 5:34:54 pm (dentsply sirona charlotte time).Warranty: dealer on-site: billable: billing accepted: billing approved by: 90 day expiration date: 04/15/2026.Remote access:tv sidexis version: server location: pdata remote location: rcu location (if applicable): practice management:tracker workstation(s): op1 scout check: troubleshooting description: image seem to be connection related office didn't have a cable so tech replaced the elastomeric strip took about 8 images back to back successfully.
|