| Model Number |
UM-986 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a pediatric surgery (laparoscopic and open procedure), the suture experienced needle and thread detachment prior to use.The issue was resolved by using a new suture.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted one suture and one needle.The needle was returned attached to the suture.The suture was returned broken 4 inches from the needle attachment point.The measurement was taken with an aluminum rule.The foil pouch was inspected and no signs of damage or abnormalities were identified.It was reported that the suture experienced needle and thread detachment prior to use.The reported issue could not be confirmed.The most likely cause could not be established from the information available.This issue can occur when exposed to the environment or clinical application.Damage during use may result in inadequate sample length to perform testing and/or deformation of the suture contributing to a loss in tensile strength.The evaluation detected an unreported condition: suture broken.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: progressive loss of tensile strength and eventual absorption of the absorbable sutures occurs by means of hydrolysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|