Catalog Number PCEE45A
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 2/11/2026 investigation summary the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that one pcee45a device was returned with no apparent damage and with no reload present.The manual override door out of position and missing; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequence firings.The bailout system was reset and then an attempt to fire the device was made, however, the device would not fire.Upon evaluation, the pcb board was noted to be non-functional.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch as part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the pcb board to be damaged.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.A manufacturing record evaluation was performed for the finished device pcee45a lot number a98f7t and no non-conformances were identified.
 
Event or Problem Description
It was reported that during an unknown surgery, could not fire.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERED PLUS COMPACT 45
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
3035526892
MDR Report Key24316085
Report Number3005075853-2026-01288
Device Sequence Number4838243
Product Code GDW
UDI-Device Identifier10705036015338
UDI-Public(01)10705036015338(17)280630(10)A98F7T(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K163454
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPCEE45A
Device Lot NumberA98F7T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/06/2026
Initial Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-