(b)(4).Date sent 2/11/2026 investigation summary the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that one pcee45a device was returned with no apparent damage and with no reload present.The manual override door out of position and missing; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequence firings.The bailout system was reset and then an attempt to fire the device was made, however, the device would not fire.Upon evaluation, the pcb board was noted to be non-functional.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch as part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the pcb board to be damaged.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.A manufacturing record evaluation was performed for the finished device pcee45a lot number a98f7t and no non-conformances were identified.
|