| Model Number |
24355 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to solventum for analysis; therefore, solventum is unable to verify the leak, identify exact location and root cause without the sample.Based on the problem description, the reported issue is most likely a heat exchanger leak.Possible causes of heat exchanger leaks include over-pressurization above 300mmhg, insertion or removal of pressurized heat exchanger from heating unit, and heat exchanger not fully inserted into the heating unit.Solventum will continue to monitor.
|
| |
|
Event or Problem Description
|
|
A user facility reported a 3m¿ ranger¿ blood/fluid warming high flow set leaked at the cassette when changing the blood bag.No injuries or medical interventions were required due to the alleged malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample was received for analysis.The reported cassette leak was confirmed by the sample; however, a root cause could not be determined.A review of the batch record showed this lot met all specifications for product release and no deviations were found that could have caused or contributed to the reported malfunction.Solventum will continue to monitor.
|
| |
|