MEDTRONIC PUERTO RICO VILLALBA SPRINT QUATTRO SECURE S MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 693565
Medical Device Problem Codes High impedance (1291); Connection Problem (2900); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/21/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Correction: h3 and h6 (annex b, c and d).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Continuation of d10: ddpb3d1 icd - implanted (b)(6) 2026.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead exhibited rising and out of range (oor) high pacing impedance.The lead remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated the impedance trend of the right ventricular pacing lead was rising.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported that the patient was brought to the catheterization lab where it was found that the right ventricular (rv) l ead pin was not past the set screw in the header of the implantable cardioverter defibrillator (icd).The lead was reinserted into the device.The lead and device remain in use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key24316637
Report Number2649622-2026-04036
Device Sequence Number4862894
Product Code LWS
UDI-Device Identifier00613994446350
UDI-Public(01)00613994446350(17)101119
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/19/2010
Device Model Number693565
Device Catalogue Number693565
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2026
Supplement Date Received by Manufacturer02/11/2026
02/11/2026
Initial Report FDA Received Date02/11/2026
Supplement Report FDA Received Date02/11/2026
02/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age78 YR
Patient SexMale
Patient Weight118 KG
-
-