MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700? CX TENACIO? PUMP; PROSTHESIS, PENIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700? CX TENACIO? PUMP; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 72404431
Medical Device Problem Codes Deflation Problem (1149); Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10: detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Incomplete inflation is acknowledged within the dfu and are not related to off label use or failure to follow instructions.The dfu provides guidance and instruction to achieve successful ams 700 implantation and to prevent inflation issues.Incomplete deflation is acknowledged within the dfu and are not related to off label use or failure to follow instructions.The dfu provides guidance and instruction to achieve successful ams 700 implantation and to prevent deflation issues.Based on the information available the cause that contributed to the reported event cannot be established.
 
Event or Problem Description
It was reported that a patient with an inflatable penile prosthesis (ipp) implant device had noted issues with inflation and deflation.It takes the patient 20 pumps to fully inflate the device and the patient cannot fully deflate the ipp device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700? CX TENACIO? PUMP
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
5086521217
MDR Report Key24316709
Report Number2124215-2026-07802
Device Sequence Number20253154
Product Code JCW
UDI-Device Identifier00191506021320
UDI-Public(01)00191506021320(17)260709(10)1100543720
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/09/2026
Device Model Number72404431
Device Catalogue Number72404431
Device Lot Number1100543720
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2026
Initial Report FDA Received Date02/11/2026
Date Device Manufactured07/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-