Model Number 24355
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to solventum for analysis.Solventum is unable to verify root cause without a sample.Based on the problem description and questionnaire response, the reported issue is most likely a heat exchanger leak.Possible causes of heat exchanger leaks include over-pressurization above 300mmhg, insertion or removal of pressurized heat exchanger from heating unit, and heat exchanger not fully inserted into the heating unit.A review of the batch record showed this lot met all specifications for product release and no deviations were found that could have caused or contributed to the reported malfunction.Solventum will continue to monitor.
 
Event or Problem Description
It was reported after setting up the warmer, fluid leakage was observed from the 3m¿ ranger¿ blood/fluid warming high flow set.No injuries or medical interventions were reported due to the alleged malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3M¿ RANGER¿ BLOOD/FLUID WARMING HIGH FLOW SET
Common Device Name
WARMER, BLOOD, NONELECTROMAGNETIC RADIATION
Manufacturer (Section D)
3M COMPANY
2510 conway ave
st. paul MN 55144
Manufacturer (Section G)
NPA DE MEXICO S. DE R.L. DE C.V.
sor juana ines de la cruz #201
50 interior 5, cd. industrial
tijuana,baja california 22444
MX   22444
Manufacturer Contact
yvonne mcmann
6203 farinon drive
san antonio, TX 78249
2103743188
MDR Report Key24317190
Report Number2110898-2026-00008
Device Sequence Number5572472
Product Code BSB
UDI-Device Identifier10707387792685
UDI-Public0110707387792685112410171727101710HX9846
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K973741
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24355
Device Catalogue Number24355
Device Lot NumberHX9846
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2026
Initial Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-