COOK INC NCIRCLE TIPLESS STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number G17520
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 - phone number: (b)(6) g4 - pma/510(k) #: exempt h3 - device evaluation anticipated but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported that the sheath of on ncircle tipless stone extractor broke prior to use during a bladder stone removal procedure.Upon opening the inner packaging of the basket, a crease was noted to be near the handle making the basket impossible to open.Upon further inspection, the sheath near the handle was discovered to be broken.The device was replaced with a new basket, and the procedure was completed successfully.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCIRCLE TIPLESS STONE EXTRACTOR
Common Device Name
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key24317196
Report Number1820334-2026-00163
Device Sequence Number5572476
Product Code FFL
UDI-Device Identifier00827002175204
UDI-Public(01)00827002175204(17)280926(10)16802041
Combination Product (Y/N)N
Initial Reporter CountryCN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG17520
Device Catalogue NumberNTSE-045065-UDH
Device Lot Number16802041
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2026
Initial Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-