| Model Number |
H74939352050 |
| Medical Device Problem Codes |
No Display/Image (1183); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a catheter issue occurred.The 70% stenosed target lesion was located in the moderately tortuous and moderately calcified distal left anterior descending artery.The opticross hd imaging catheter was selected for intravascular ultrasound examination.During the procedure, blackout issues were encountered with the ivus catheter, which disrupted imaging and created challenges during the procedure.Despite repeated user attempts to eliminate air bubbles through multiple flushing, the issue persisted.The procedure was completed with another of the same device.The patients condition was good, and no complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a catheter issue occurred.The 70% stenosed target lesion was located in the moderately tortuous and moderately calcified distal left anterior descending artery.The opticross hd imaging catheter was selected for intravascular ultrasound examination.During the procedure, blackout issues were encountered with the ivus catheter, which disrupted imaging and created challenges during the procedure.Despite repeated user attempts to eliminate air bubbles through multiple flushing, the issue persisted.The procedure was completed with another of the same device.The patients condition was good, and no complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: device analysis findings: the device was returned for analysis.Visual inspection revealed no issues, and the device appeared to be in good condition.Impedance testing revealed no electrical issues with the device and a good square image appeared in the ilab system during image characterization testing.During the flush test, the device flushed normally when the telescope was fully retracted and fully advanced.No signs of leak were observed.Device history record (dhr) review: it was confirmed that this device met manufacturing specification prior to distribution and there no manufacturing deviations which could have contributed to the reported event.Investigation conclusion: this investigation has been assigned an investigation conclusion code of device problem excluded following a review of the returned device, reported event details, available information, and product record review.In this case, the device performed as intended during the analysis, and no damage was found that could have contributed to the reported allegation.
|
| |
|
Search Alerts/Recalls
|