| Model Number |
8422400 |
| Medical Device Problem Codes |
No Display/Image (1183); Display or Visual Feedback Problem (1184); Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during turning the monitor to one side it goes blank and when turn it to the other the side it will come back on.No patient health consequences have been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Event or Problem Description
|
|
It was reported that during turning the monitor to one side it goes blank and when turn it to the other the side it will come back on.No patient health consequences have been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The evita vn800 device was analyzed via provided information incl.The secured log files.According to the log, it could be identified that the error messages ¿device failure (14)¿ and the subsequent ¿alarm system failed¿ were issued at the reported time.Both alarms resulted from a temporary impairment of the internal hdbaset communication between the ecd and the ventilation unit.In such cases, display functions may be affected and the screen may turn black.In case of an internal failure of the display unit (ecd) and/or an impaired internal hdbaset communication between the ventilation unit and the ecd the alarm message "device failure (14)" will be raised.The display functions and operability of the ecd might be affected in case of an active "device failure (14)" alarm event depending on the technical root cause.If the internal hdbaset communication fails completely, the secondary alarm (piezo speaker) of the ventilation unit will sound.Safety relevant parameters as fio2, minute volume and airway pressure are displayed on the supplemental oled-display wherein the user can observe the ongoing ventilation.The device reacted as specified and was checked onsite according to manufacturer specs.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|
| |
|