This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Customer returned pump for an alleged cracked battery compartment on event date 03-jun-2025 pump received with cracked belt clip slot, broken battery tube threads, stained end cap sticker, stained address/serial number label and cracked reservoir tube lip.The pump passed the displacement test and the test p-cap/reservoir does lock into place.Cosmetic damage was confirmed at cracked case (battery tube).Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|