MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Medical Device Problem Code Failure to Fold (1255)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported during cataract phacoemulsification and intraocular lens (iol) implant procedure, the injector does not properly load the lens, preventing its implantation during loading.The procedure was completed using another lens of the same type with no patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information provided in sections d.9., h.3., h.6., and h.11.The device and the lens were returned loose in the lens carton.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was dried in the device.The plunger was fully advanced with the plunger tip outside the nozzle tip.No issues or abnormalities observed.The lens was returned outside the device loose in the carton with the device.Viscoelastic was dried on the lens.The lens was cleaned with klrp for further evaluation.Both haptics were pliable when manipulated.No issues observed.A fold inspection was conducted using folding tweezers.The lens folded and unfolded with no abnormalities observed.A dimensional (plan view) inspection was conducted and the lens met specification per the approved (plan view) template.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Associated viscoelastic was not provided.It is unknown if a qualified product was used.No problems were observed with the returned device or lens.Associated viscoelastic was not provided.It is unknown if a qualified product was used.The instructions for use (ifu) instruct: during device preparation and implantation of the company iq iol with the company preloaded delivery system, a company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Only use an alcon ovd qualified for use with the company pre-loaded delivery system that has been allowed to come to the operating room temperature.The ifu instructs: take at least 7 seconds to gently fold the intraocular lens (iol) by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the ¿fill-to¿ line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.After the lens has been advanced to the ¿fill-to¿ line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.No part of the lens should exit the nozzle prior to insertion through the incision.Once the lens is in position at the ¿fill-to¿ line, it should be implanted within 3 minutes.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome.The manufacturer internal reference number is: 2026-05804.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24320684
Report Number1119421-2026-00292
Device Sequence Number21310263
Product Code KYB
UDI-Device Identifier00380652394819
UDI-Public(01)00380652394819(17)280728(11)250730(10)16085563
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAU00T0
Device Lot Number16085563
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/20/2026
Supplement Date Received by Manufacturer03/09/2026
Initial Report FDA Received Date02/12/2026
Supplement Report FDA Received Date04/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-