| Model Number |
0684-00-0607 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during insertion of the transray plus 35cc catheter shaft kinked and became unusable.There is no information about patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to characterization limit e1 initial reporter: (b)(6).Due to characterization limit e1 event site name: (b)(6).The device was not returned and could not be evaluated.It was discarded by the user.We are unable to confirm the reported event.If new information becomes available, a supplemental report will be submitted.Complaint record id (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields- b4, e1, g3, g6, h1, h2, h6 (type of investigation, investigation findings, component codes, investigation conclusions), h11.No decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Since the product was not returned, we were unable to evaluate the device.Therefore, the reported failure cannot be confirmed and cannot determine a root cause.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields - b4, g3, g6, h2, h6 (health effect clinical code ), h11 corrected field - b5, e3.
|
| |
|
Event or Problem Description
|
|
It was reported that during insertion of the transray plus 35cc catheter shaft kinked and became unusable.There is no harm or injury.
|
| |
|