| Catalog Number |
ES2916 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no non-conforming reports or capas that were confirmed in veeva to be associated.The device was not returned for evaluation.Without the benefit of analyzing the device, coloplast cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information, during intraoperative preparation of a 16 cm touch scrotal device, a persistent air infiltration issue was observed.Despite multiple preparation attempts, air continued to return to the pump mechanism.The device was determined to have a suspected manufacturing defect [leak] and was replaced with a new unit.
|
| |
|
Event or Problem Description
|
|
According to the available information, during intraoperative preparation of a 16 cm touch scrotal device, a persistent air infiltration issue was observed.Despite multiple preparation attempts, air continued to return to the pump mechanism.The device was determined to have a suspected manufacturing defect [leak] and was replaced with a new unit.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump and cylinders 1 and 2 were received for evaluation.No functional abnormalities were noted with the pump.No functional abnormalities were noted with either cylinder 1 or cylinder 2.The information received indicated the device had air infiltration issues, but because no functional abnormalities were noted with the returned components, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|