| Catalog Number |
BSTPULR6250SL |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In this event it is reported that a pro-taper ultimate slider 25mm x6 file broke during use.The broken part could not be retrieved from the tooth.Additional treatment will be needed, including surgery.Further information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
While no injury was reported, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation results for this complaint are involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.No link can be established between breakage issue and information found during dhr review (batch #1921317).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique, overuse (number of uses not communicated), excessive wear, patient condition and behavior during treatment or material issue (no analysis of the broken fragments possible).In addition, there are further circumstances on the conditions in dentistry (used disinfectants, training/knowledge status), or any other environmental conditions, which are unknown to us and may also have an impact on the reported failure mode.
|
| |
|
Search Alerts/Recalls
|