|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- sensor images have artifacts.Sensor was replaced.
|
|
While the customer was using a part of an intraoral sensor system, schick33 s2 starter kit/3.0 interface 6', they allege experiencing image quality issues when an x-ray image was taken and an additional image was required.No injury was reported from the alleged event.
|