While the customer was using a part of an intraoral sensor system, schick33 s2 starter kit/3.0 interface 6', they allege there was an x-ray image transmission issue where the x-ray image was not captured after radiation has occurred (transmission failure).No injury was reported from the alleged event.
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- solution description 01/19/2026; 12:23:08; (b)(4).Solution description: push the update to the remote from 121 to 140.Issues were with software configuration and device recognition.If the issues device is working now if the problem follows the device then replace the remote.
|