| Model Number |
2110 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: the device was received for evaluation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.Visual and functional tests were performed, which found the device's upstream occlusion (uso) seal was buckled.The uso seal was replaced to fix the issue.
|
| |
|
Event or Problem Description
|
|
The device was returned as it was reported that when the pump was plugged in, the charging port failed to recognize the ac adapter.The results of the investigation found that the device's upstream occlusion (uso) seal was buckled.There was no patient involvement.
|
| |
|