• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH IMN INSTRUMENTS - DISTAL TARGETING ARM FEMUR ANTEGRADE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH IMN INSTRUMENTS - DISTAL TARGETING ARM FEMUR ANTEGRADE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 23533106
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Event or Problem Description
As reported: "due to insufficient adjustment of the adjusting device, the most distal advance locking screw of the two distal transverse stops was inserted into the bone without passing through the nail.This was discovered after wound closure.Reinsertion of the screw outside the nail was not performed, and the plan is to observe the situation and take action as appropriate.".
 
Additional Manufacturer Narrative
Corrections: please refer to section h6 (health impact code was revised).The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
As reported: "due to insufficient adjustment of the adjusting device, the most distal advance locking screw of the two distal transverse stops was inserted into the bone without passing through the nail.This was discovered after wound closure.Reinsertion of the screw outside the nail was not performed, and the plan is to observe the situation and take action as appropriate.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMN INSTRUMENTS - DISTAL TARGETING ARM FEMUR ANTEGRADE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA GMBH
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key24328429
Report Number0008031020-2026-00162
Device Sequence Number10005268
Product Code HSB
UDI-Device Identifier07613327357066
UDI-Public(01)07613327357066(11)191211(10)K08DDF3
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172774
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number23533106
Device Lot NumberK08DDF3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/16/2026
Supplement Date Received by Manufacturer03/04/2026
Initial Report FDA Received Date02/12/2026
Supplement Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-