RESPIRONICS, INC. SYSTEM ONE HEATED HUMIDIFIER; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number IN6H
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
A device was returned to third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and found evidence of foam degradation.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service center.There were no error codes found.The third-party service center concludes that there was a visible foam degradation.
 
Additional Manufacturer Narrative
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM ONE HEATED HUMIDIFIER
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key24329899
Report Number2518422-2026-101023
Device Sequence Number4876749
Product Code BZD
UDI-Device Identifier00606959014244
UDI-Public(01)00606959014244
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberIN6H
Device Catalogue NumberIN6H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2024
Initial Report FDA Received Date02/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-