Model Number EGIA45CTAV
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Tissue Breakdown (2681)
Date of Event 01/26/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, intra-operatively, immediately after stapling, a small bleeding was observed at the distal portion of the staple line.The physician promptly clamped the area with a surgical clamp to prevent further bleeding, followed by conversion to an open procedure.Sutures were placed at the bleeding site.A new stapling was not possible, as the remaining stump was too small to accommodate another stapler application.The surgical time was extended.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the staple line was incomplete.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24330184
Report Number1219930-2026-00897
Device Sequence Number14707095
Product Code GDW
UDI-Device Identifier10884523003192
UDI-Public(01)10884523003192(17)280131(10)N3B0155Y
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEGIA45CTAV
Device Catalogue NumberEGIA45CTAV
Device Lot NumberN3B0155Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2026
Supplement Date Received by Manufacturer05/15/2026
Initial Report FDA Received Date02/13/2026
Supplement Report FDA Received Date05/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
Patient Weight66 KG
-
-