The device with the lens was returned.The plunger lock and lens stop were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was at the trailing optic edge.The lens was advanced to mid-nozzle.The trailing haptic was folded on the optic.The leading haptic was extended.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A non-qualified viscoelastic was indicated.No problems were observed with the returned device.The plunger, lens and haptic positions were acceptable.The leading haptic was straight.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the instructions for use (ifu).The customer indicated the use of a non-qualified viscoelastic which could be a contributing factor.A straight leading haptic may occur: due to the normal folding variations as indicated in the ifu diagrams.If the initial plunger advancement is faster than the recommended target, the leading haptic may be forced past the internal folding feature.If there is excessive delay between the device preparation and subsequent delivery the leading haptic may start to unfold.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ophthalmic viscosurgical device (ovd)s may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (greater than 23 degrees celsius or 73 degrees fahrenheit) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement or contribute to incorrect haptic folding.Any of the above listed causes alone, or in combination, may create the reported event.The ifu instructs: insert the tip of the cartridge through the incision.Position the tip at the anterior capsule opening.If the leading haptic is looped or straight in front of the optic, rotate the nozzle clockwise as needed to be bevel left to facilitate placement of the leading haptic into its normal orientation in the bag.Slowly advance the lens until the optic is exiting the nozzle.Rotate the cartridge counter-clockwise towards bevel right as needed to place the lens anterior side up.The manufacturer internal reference number is: (b)(4).
|