• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUTONIX, INC LUTONIX 035 DRUG COATED PTA DILATION CATHETER; DRUG COATED BALLOON PTA CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUTONIX, INC LUTONIX 035 DRUG COATED PTA DILATION CATHETER; DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9004
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent a percutaneous transluminal angioplasty (pta) procedure using the lutonix 035 drug coated balloon catheter.During the procedure, it was noticed that the balloon catheter allegedly ruptured.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.One video was received and reviewed.The video shows one healthcare professional holding the balloon in an inflated condition while another inflates the balloon using a syringe/inflation device filled with fluid.During this process, fluid leakage from the balloon was observed.No additional anomalies were noted in the video.Based on the video review, the investigation is confirmed for the reported balloon rupture as a fluid leakage from the balloon was noted.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h6 (method, result).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent a percutaneous transluminal angioplasty (pta) procedure using the lutonix 035 drug coated balloon catheter.During the procedure, it was noticed that the balloon catheter allegedly ruptured.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: the initial mdr was inadvertently submitted with a g3 date of 01/18/2026.The correct g3 date is 02/11/2026.Manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.One video was received and reviewed.The video shows one healthcare professional holding the balloon in an inflated condition while another inflates the balloon using a syringe/inflation device filled with fluid.During this process, fluid leakage from the balloon was observed.No additional anomalies were noted in the video.Based on the video review, the investigation is confirmed for the reported balloon rupture as a fluid leakage from the balloon was noted.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, g2, g3.Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent a percutaneous transluminal angioplasty (pta) procedure using the lutonix 035 drug coated balloon catheter.During the procedure, it was noticed that the balloon catheter allegedly ruptured.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED PTA DILATION CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
LUTONIX, INC
9409 science center dr
new hope 55428
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key24331455
Report Number3006513822-2026-00008
Device Sequence Number9791406
Product Code ONU
UDI-Device Identifier00801741163418
UDI-Public(01)00801741163418(11)240809(17)270702(10)GFJR1572
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9004
Device Catalogue NumberLX3513052205F
Device Lot NumberGFJR1572
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2026
Supplement Date Received by Manufacturer03/06/2026
03/06/2026
Initial Report FDA Received Date02/13/2026
Supplement Report FDA Received Date03/11/2026
03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-