| Model Number |
9004 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a percutaneous transluminal angioplasty (pta) procedure using the lutonix 035 drug coated balloon catheter.During the procedure, it was noticed that the balloon catheter allegedly ruptured.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.One video was received and reviewed.The video shows one healthcare professional holding the balloon in an inflated condition while another inflates the balloon using a syringe/inflation device filled with fluid.During this process, fluid leakage from the balloon was observed.No additional anomalies were noted in the video.Based on the video review, the investigation is confirmed for the reported balloon rupture as a fluid leakage from the balloon was noted.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h6 (method, result).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a percutaneous transluminal angioplasty (pta) procedure using the lutonix 035 drug coated balloon catheter.During the procedure, it was noticed that the balloon catheter allegedly ruptured.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the initial mdr was inadvertently submitted with a g3 date of 01/18/2026.The correct g3 date is 02/11/2026.Manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.One video was received and reviewed.The video shows one healthcare professional holding the balloon in an inflated condition while another inflates the balloon using a syringe/inflation device filled with fluid.During this process, fluid leakage from the balloon was observed.No additional anomalies were noted in the video.Based on the video review, the investigation is confirmed for the reported balloon rupture as a fluid leakage from the balloon was noted.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, g2, g3.Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a percutaneous transluminal angioplasty (pta) procedure using the lutonix 035 drug coated balloon catheter.During the procedure, it was noticed that the balloon catheter allegedly ruptured.There was no reported patient injury.
|
| |
|