| Model Number |
WA50023B |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the rigid video laparoscope's light was very dim and the scope appeared denied.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h3, h6 and h11 the device was returned to olympus for inspection and the reported failure for the scope being denied/connection problem could not be confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: fiber breakage and dents on distal edge.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|