MENTOR TEXAS SMOOTH MOD HIGH BOOST, 635 CC; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SMHB635 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/28/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 43-year-old caucasian female patient underwent revision breast augmentation with smooth mod high boost, 635 cc and experienced capsular contracture; baker grade iii on her right side postoperatively; diagnosed in the office.The patient has consulted with the healthcare professional.As a result, the patient is scheduled for surgery on (b)(6) 2026.Mentor is aware that the event date (january 28, 2025) is prior to the manufacturing date (april 20, 2025) of reported lot number 2069457.If additional information becomes available, it will be updated and supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: right capsular contracture; baker grade iii.D6b explantation date: (b)(6) 2026.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On april 14, 2026, the mentor failure analysis lab received the device for evaluation.Clarification has been requested for the date of explant.Initially (b)(6) 2026 was provided and later with the device, date of explant was provided as (b)(6) 2026.Supplemental report will be submitted upon receipt of any further information.On april 29, 2026, device evaluation was completed as follows: mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The contralateral device was also received (lot number 2055310).The patient did not report any issue with this device.Therefore, no further analysis of the contralateral device is required.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|