| Catalog Number |
205513000 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that it did not happen in surgery, spd found debris coming out of the locking mechanism and cannot be cleaned.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: according with information received, it was reported that it did not happen in surgery, spd found debris coming out of the locking mechanism and cannot be cleaned.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned device found foreign substance of the s/c trial handle angled.The device exhibits an overall worn appearance consistent with normal and constant usage.Per ifu-0902-00-721, care should be taken to remove any debris, tissue, or bone fragments that may collect on the instrument.Instruments with cannulas, moving parts, or spring mechanisms require additional cleaning steps to ensure proper removal of all trapped debris.In addition to other important steps to follow, the ifu instructs the use of a soft bristle brush to remove all traces of blood and debris, paying close attention to threads, crevices, seams, and any hard-to-reach areas of the device features.If the instrument has sliding mechanisms or hinged joints, actuate the area to free any trapped blood and debris.A dimensional inspection was not performed since it was not applicable to the complaint condition.A manual functional test performed identified that one of the ball plungers was not retracting properly.The reported condition of foreign substance can be confirmed.The overall complaint was confirmed as the observed condition of the s/c trial handle angled would have contribute to the complained device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.Corrected: d4 (primary udi number).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|