MENTOR TEXAS MENTOR MEMORYGEL BOOST BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SHPB480 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Hematoma (1884)
|
| Date of Event |
01/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 33-year-old female patient underwent a primary breast augmentation with a 480cc mentor memorygel boost breast implant and experienced capsular contracture (baker grade 3) on the right side post-operatively, which was diagnosed with an office visit.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
|
| |
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: no information regarding explantation or an expected explantation date has been received.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On february 17, 2026, mentor received additional information reporting that the patient was to undergo device explantation on (b)(6) 2026.D6b.Explantation date: (b)(6) 2026.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Reason for device explant and/or reoperation: capsular contracture.On february 19, 2026, mentor received additional information regarding the patient's experience.It was reported that the patient has been taking singulair but has not noticed any changes.The breast was reported to be sitting higher and be firmer.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On march 27, 2026, mentor received additional information reporting that the patient's replacement device was a 480cc mentor memorygel boost breast implant.On april 07, 2026, mentor received additional information regarding the patient's event.It was reported that the patient's postoperative diagnosis, the patient also experienced an implant site hematoma.The also experienced hardness and pain.Code e0505 for hematoma has been added in section h6-health effect - clinical code to document this additional information.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On april 17, 2026, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On may 07, 2026, the evaluation for the device was completed.Device evaluation summary:
during visual evaluation bubbles within the gel were observed on the returned device.Leak testing was performed, in accordance with mentor procedures, and no areas of gel exposure were detected during the analysis.The shell wall of the implant is permeable to air, and trapped air can form bubbles with changes in pressure or temperature.When left by themselves with no changes in temperature or pressure, bubbles usually dissipate over time unless trapped by cured gel.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.Hematoma is a collection of blood within the space around the implant.Symptoms from hematoma may include swelling, pain, and bruising.If a hematoma occurs, it will most likely occur soon after surgery; however, it can occur at anytime, such as after an injury to the breast.Small hematomas may be absorbed by the body, while others may require surgery for drainage.Hematoma is a known complication associated with this procedure and is referenced in our product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number:
(b)(4).
|
| |
|