| Model Number |
E142 |
| Medical Device Problem Codes |
High impedance (1291); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) was explanted due to a product performance issue.This pacemaker was successfully explanted and a new lead implanted instead.No additional adverse patient effects were reported outside the revision procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up report 1 (additional information): this report is being submitted to update field b5: describe event or problem, section b and field h6: device codes, section h.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) was explanted due to a product performance issue.This pacemaker was successfully explanted and a new lead implanted instead.No additional adverse patient effects were reported outside the revision procedure.Additional information was received that this icd was explanted recorded high out of range shock impedance measurements on the right ventricular (rv) channel and the device was included on the revision procedure as it was nearing elective replacement indicator (eri).No additional adverse patient effects were reported outside the revision procedure.
|
| |
|