MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ESSENTIO? SR; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ESSENTIO? SR; PACEMAKER Back to Search Results
Model Number L100
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) reverted into safety mode.This device was confirmed to be in safety mode and device replacement was recommended.No adverse patient effects were reported.This crt-p remains in service.
 
Additional Manufacturer Narrative
Follow up report 1 (device evaluation): the complaint device was not returned to boston scientific; therefore, product analysis could not be performed.Conclusion code was updated to "cause not established" because the reason for the alleged observation(s) could not be determined.Product record reviews did not identify a product quality issue and analysis of available information did not identify a specific complaint cause.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of device displays error message was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.At this point, it can be concluded that unknown factors most probably contributed to the event.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) reverted into safety mode.This device was confirmed to be in safety mode and device replacement was recommended.No adverse patient effects were reported.This crt-p remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO? SR
Common Device Name
PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24335040
Report Number2124215-2026-08299
Device Sequence Number18578015
Product Code LWP
UDI-Device Identifier00802526558900
UDI-Public(01)00802526558900(17)190407
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
N970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/07/2019
Device Model NumberL100
Device Catalogue NumberL100
Device Lot Number717355
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2026
Supplement Date Received by Manufacturer04/27/2026
Initial Report FDA Received Date02/13/2026
Supplement Report FDA Received Date04/28/2026
Date Device Manufactured06/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-