| Model Number |
9-ASD-MF-035 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 35mm amplatzer cribriform was chosen for implant using an 9f amplatzer trevisio delivery system.During the procedure, while deploying, it was noted that the occluder did not conform to its desired shape but it deformed like "invert martini glass".The device was recaptured trice and had to retrieve outside of the patient's body to examine.The occluder was massaged in a saline bath to help get back to its desired shape and then it was re prepared.When attempted again, the same issue was observed.The procedure was completed using a larger device.The deformed device was never released from the delivery cable.There was no interaction with the cardiac structures during deployment and no angulation/kink were noticed in the delivery system.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.There were no complaints associated with any other devices from the lot.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Search Alerts/Recalls
|