| Catalog Number |
100008634 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- solution description 01/23/2026 16:35:47 rebecca buck (rbuck) troubleshooting description: artifacts in sensor.They just started using the sensor and noticed the artifact in a corner.They noticed after a few images on one patient.Attached.She took a test on herself and i posted it.The artifact does not appear in blank area.Attached.I did a session on op2 and see firmware version and updated.
|
| |
|
Event or Problem Description
|
|
While the customer was using a part of an intraoral sensor system, schick33 s2 starter kit/3.0 interface 9', they allege experiencing image quality issues when an x-ray image was taken and an additional image was required.No injury was reported from the alleged event.
|
| |
|
Additional Manufacturer Narrative
|
|
This is to correct and remove the codes that were initially reported - removing codes for: type of investigation : 4118.Investigation findings code: 3233.Investigation conclusions : 11.The correct codes for this complaint are: type of investigation : 10.Investigation findings code: 4214.Investigation conclusions : 4318.This is a follow up report to correct this code.
|
| |
|
Search Alerts/Recalls
|