Catalog Number BG8795U
Medical Device Problem Codes Product Quality Problem (1506); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No:(b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section e.1: the initial reporter phone is not available / reported.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (8943735) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the limited information available and without the product available for analysis, the reported issue documented in the complaint could not be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and / or device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered later.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during a mechanical thrombectomy procedure targeting an occlusion in the common carotid artery (cca), a combination of a ¿half stiff wire and a jb2 catheter was advanced into the aorta.Because the vessel was markedly tortuous, the inner catheter was kept positioned in the aorta on standby while the wire was advanced first.¿ the 95cm emboguard 87 balloon guide catheter (bgc) (bg8795u / 8943735) was then advanced along the wire from the aortic arch to the cca, ¿however, angiographic image was somewhat poor and the emboguard was confirmed to be kinked partially.Then the cereglide was advanced without issues and used afterwards.The procedure was successfully completed.¿ continuous flush was maintained during the procedure.There was no negative impact to the patient.On 12-feb-2026, additional information was received.The information indicated that a jb2 catheter was also used.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to include the additional event information received on 23-feb-2026.[additional information]: on 23-feb-2026, additional information was received.Per the information, the surgical access point was femoral.An introducer sheath was used but the brand was unknown.There was also the use of a peel-away sheath.The emboguard bgc was inflated one time during the procedure.The information indicated that the vessel was a type 3 and tortuosity was relatively severe.There was no ussie with the marker bands.The information confirmed there was no clinically significant delay in the procedure.Updated sections: b.4, g.3, g.6, h.2, and h.11.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOGUARD
Common Device Name
CATHETER, PERCUTANEOUS, NEUROVASCULATURE
Manufacturer (Section D)
NEURAVI LTD.
block 3
ballybritt business park
galway H91 K 5YD
EI  H91 K5YD
Manufacturer (Section G)
LAKE REGION MEDICAL GALWAY
parkmore business park west
county galway H91 C K22
EI   H91 CK22
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
949788-2078
MDR Report Key24338017
Report Number3011370111-2026-00018
Device Sequence Number4914448
Product Code QJP
UDI-Device Identifier10886704082743
UDI-Public(01)10886704082743(11)240229(17)260831(10)8943735
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K212340
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Catalogue NumberBG8795U
Device Lot Number8943735
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2026
Supplement Date Received by Manufacturer02/23/2026
Initial Report FDA Received Date02/13/2026
Supplement Report FDA Received Date02/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
JB2 CATHETER (DOTMED).; TREVO TRAK ® 21 MICROCATHETER (STRYKER).
Patient SexUnknown
-
-