| Catalog Number |
BG8795U |
| Medical Device Problem Codes |
Product Quality Problem (1506); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No:(b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section e.1: the initial reporter phone is not available / reported.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (8943735) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the limited information available and without the product available for analysis, the reported issue documented in the complaint could not be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and / or device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered later.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
The healthcare professional reported that during a mechanical thrombectomy procedure targeting an occlusion in the common carotid artery (cca), a combination of a ¿half stiff wire and a jb2 catheter was advanced into the aorta.Because the vessel was markedly tortuous, the inner catheter was kept positioned in the aorta on standby while the wire was advanced first.¿ the 95cm emboguard 87 balloon guide catheter (bgc) (bg8795u / 8943735) was then advanced along the wire from the aortic arch to the cca, ¿however, angiographic image was somewhat poor and the emboguard was confirmed to be kinked partially.Then the cereglide was advanced without issues and used afterwards.The procedure was successfully completed.¿ continuous flush was maintained during the procedure.There was no negative impact to the patient.On 12-feb-2026, additional information was received.The information indicated that a jb2 catheter was also used.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to include the additional event information received on 23-feb-2026.[additional information]: on 23-feb-2026, additional information was received.Per the information, the surgical access point was femoral.An introducer sheath was used but the brand was unknown.There was also the use of a peel-away sheath.The emboguard bgc was inflated one time during the procedure.The information indicated that the vessel was a type 3 and tortuosity was relatively severe.There was no ussie with the marker bands.The information confirmed there was no clinically significant delay in the procedure.Updated sections: b.4, g.3, g.6, h.2, and h.11.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Search Alerts/Recalls
|
|
|