This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.
|
|
It was reported that during service and evaluation, it was determined that the motor device was loose, had a damaged flex circuit, was making excessive noise, had vibration, and cosmetic damage.It was noted that the device connector was loose, the connector red dot was fading, the housing pin was loose, the set screw was loose, there was electrical leakage, abnormal noise, and vibration.It was further determined that the device failed pretest for visual assessment, loctite assessment, no short circuit between phases and connector body, no short circuit between hall sensors and connector body, no short circuit between 5vdc line and connector body, and noise assessment.It was noted in the service order that the device had an undetermined malfunction.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported; however, it was reported that the event occurred in 2026.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|