• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ PID; GRAM POSITIVE IDENTIFICATION PANEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ PID; GRAM POSITIVE IDENTIFICATION PANEL Back to Search Results
Catalog Number 448008
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using the bd phoenix¿ pid a patient isolate (staphylococcus epidermidis) was misidentified as staphylococcus aureus or staphylococcus capitis.The user verified the final result by performing repeat testing using a reference laboratory.No health impact or consequence reported.This is report 5 of 6.
 
Event or Problem Description
It was reported while using the bd phoenix¿ pid a patient isolate (staphylococcus epidermidis) was misidentified as staphylococcus aureus or staphylococcus capitis.The user verified the final result by performing repeat testing using a reference laboratory.No health impact or consequence reported.This is report 5 of 6.
 
Additional Manufacturer Narrative
Investigation summary: this complaint is for misidentification when using phoenix panel pid (catalog number 448008) batch number 5254496.The customer returned phoenix generated lab reports showing staphylococcus aureus and staphylococcus epidermidis identifications for the investigation.The batch history record was satisfactory, and no quality notifications were generated during manufacturing and inspection.Complaint trending was performed and no trends were identified associated with this defect.Bd will continue to monitor for trends and take action as necessary.Please continue to communicate any additional concerns.To investigate, panels of the complaint batch and control panels from the same material but different batch were tested using in house isolate s.Epidermidis pos 3644 on a phoenix m50 machine and evaluated for identification results.The panels tested identified their inoculated isolates correctly, this complaint is not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ PID
Common Device Name
GRAM POSITIVE IDENTIFICATION PANEL
Manufacturer (Section D)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key24338109
Report Number1119779-2026-00140
Device Sequence Number9790656
Product Code LQL
UDI-Device Identifier00382904480080
UDI-Public(01)00382904480080(10)5254496
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/20/2026
Device Catalogue Number448008
Device Lot Number5254496
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2026
Supplement Date Received by Manufacturer03/02/2026
Initial Report FDA Received Date02/13/2026
Supplement Report FDA Received Date03/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/28/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-