| Model Number |
WA22306D |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: (b)(6).The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the resection electrode tip section was bent upon receipt of the device.There wee no reports of patient involvement.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the completed investigation, the root cause of the reported malfunction could not be definitively determined.However, the issue is likely attributable to component damage.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|