The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the segment fluid damage, the control body fluid damage, the lg connector fluid damage, the control body corroded, the ccd drive pcb corroded, the lg connector corroded, and the lg connector leak; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "(hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|