• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number CM-813
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a cesarian section, while suturing, when pulling the suture through the tissue, the suture detached from the needle.A total of six sutures had the same issue.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant products: cm-813, cm-813 biosyn* 1 vio 75cm gs21 x36 - (lot#: d5e4035y); cm-813, cm-813 biosyn* 1 vio 75cm gs21 x36 - (lot#: d5e4035y); cm-813, cm-813 biosyn* 1 vio 75cm gs21 x36 - (lot#: d5e4035y); cm-813, cm-813 biosyn* 1 vio 75cm gs21 x36 - (lot#: d5e4035y); cm-813, cm-813 biosyn* 1 vio 75cm gs21 x36 - (lot#: d5e4035y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.Eighty-four representative samples were returned for analysis.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that when pulling the suture through the tissue, the suture detached from the needle.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24342379
Report Number9612501-2026-00381
Device Sequence Number9541372
Product Code GAM
UDI-Device Identifier10884521032019
UDI-Public(01)10884521032019(17)300430(10)D5E4035Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCM-813
Device Catalogue NumberCM-813
Device Lot NumberD5E4035Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2026
Supplement Date Received by Manufacturer05/15/2026
Initial Report FDA Received Date02/16/2026
Supplement Report FDA Received Date05/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Patient SexUnknown
-
-