MAUDE Adverse Event Report: PERFUSION SYSTEMS PENDITURE LAA EXCLUSION SYSTEM; LEFT ATRIAL APPENDAGE CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS PENDITURE LAA EXCLUSION SYSTEM; LEFT ATRIAL APPENDAGE CLIP, IMPLANTABLE Back to Search Results
Model Number LAAC40
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the penditure clip failed to stay on the left atrial appendage during the procedure.The use of the device was unspecified.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Additional information b5.The clip was not used.Medtronic received additional information that movement/slippage was more than 10mm.The surgeon had to reposition as well as recapture a few times.The surgeon does not use a sizer for our clip or any clip for that matter.This was a coronary artery bypass grafting (cabg) procedure.The surgeon did not use another clip, the patient left the or clipless.The clip was deployed on a beating heart with high left atrial pressures, nothing special about left atrial appendage (laa) anatomy.Inferior approach was used.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Device evaluation summary: visual inspection shows no outward signs of any damage.The clip dimensions were measured to be as follows.Clip length was measured to be 1.57 inches, the specification is 1.59 +/-.04 inches.The jaw teeth height was measured to be 0.019 and 0.019 inches, the specification is 0.020 +/- 0.003 inches.The shark fins operate as expected.The clamp force was measured to be 1.666 lbs., the specification is 1.9 +/- 0.6 lbs.Reason for the return was un determined.Additional information h6.2 eval code method (fdm/annex b): this field was updated.Additional information h6.3 eval code result (fdr/annex c): this field was updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENDITURE LAA EXCLUSION SYSTEM
Common Device Name
LEFT ATRIAL APPENDAGE CLIP, IMPLANTABLE
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key24342783
Report Number2184009-2026-00251
Device Sequence Number9523824
Product Code PZX
UDI-Device Identifier00763000791452
UDI-Public(01)00763000791452(17)280902(10)T5J004X(20)01
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K232295
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLAAC40
Device Catalogue NumberLAAC40
Device Lot NumberT5J004X
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2026
Supplement Date Received by Manufacturer02/26/2026
03/18/2026
Initial Report FDA Received Date02/16/2026
Supplement Report FDA Received Date03/02/2026
04/13/2026
Date Device Manufactured09/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
-
-