Catalog Number 8065751761
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A customer reported that an ophthalmic handpiece had a failure of phacoemulsification function during surgery.The surgery was completed by switching to another handpiece on the same day.There was no patient harm.
 
Additional Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.A phaco handpiece (hp) was returned for testing on this investigation.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation with the same event code.A visual assessment of the returned sample revealed cable herniation, connector discoloration, and a missing red dot.The returned sample was connected to a calibrated system.The handpiece was recognized successfully.The handpiece tuned successfully and completed a five-minute burn-in test with the system set at 100% ultrasonic and torsional power.The handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements which found the handpiece to meet manufacturer specifications the handpiece was found to meet specifications; therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, CENTURION OZIL HANDPIECE
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24342984
Report Number2028159-2026-00178
Device Sequence Number18522623
Product Code HQC
UDI-Device Identifier00380657517619
UDI-Public(01)00380657517619(17)350407(11)250409(10)17C3X6
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K121555
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065751761
Device Lot Number17C3X6
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2026
Supplement Date Received by Manufacturer04/29/2026
Initial Report FDA Received Date02/16/2026
Supplement Report FDA Received Date05/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-