ZIMVIE US CORP LLC IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number TSVH10
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the implant was removed due a fracture at the collar at tooth site #15.No other impact on the patient reported.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4) a4: patient weight unknown / not provided.G4: premarket identification k011028/k013227.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report, b5: describe event or problem, g3: date received by manufacturer, g6: type of report, h1: type of reportable event, h2: follow up type, h3: device evaluated by manufacturer, h6: adverse event problem, h11: additional narrative.Zimvie received one (1) tsvh10, (impl tapered scr-v ha 3.7 mm 3.5mm 10mm) for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, apparent bone/tissue attached to external threads.Implant fracture identified at the collar.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number 1244105.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1244105 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿fracture implant¿ based on the investigation and risk management file review, the most likely root causes determined from the investigation are parafunctional habits/patient factors, or customer error.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24343189
Report Number0001038806-2026-00760
Device Sequence Number21323765
Product Code DZE
UDI-Device Identifier00889024019560
UDI-Public(01)00889024019560(17)260805(10)1244105
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K061410
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 04/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/05/2026
Device Catalogue NumberTSVH10
Device Lot Number1244105
Was Device Available for Evaluation? No
Date Returned to Manufacturer02/05/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/16/2026
Supplement Report FDA Received Date04/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-