It was reported that, during a second stage clinical procedure evaluation, exposure of the cover screw at implant #13 was attempted.A zimmer driver was used to engage the cover screw.An attempt was made to remove the cover screw and the entire implant was explanted.No osseaus on the surface of the implant was noted and a lack of integration was noted.After implant removal, the site was patched with bg and covered to preserve the site for a future implant replacement.
|
|
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.A2: age and date of birth unknown / not provided.A4: weight unknown / not provided.E1: email address unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|