|
|
| Catalog Number |
72201769 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H11:this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a patellofemoral ligament repair procedure, the front end of the biosure ha anchor broke while being driven in by squeezing.The pieces were removed from inside the patient using tweezers and suction.The procedure was resumed, after a 10 minute delay, using an equivalent s+n back-up inserted into the original bone hole.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: the reported device was received for evaluation.A visual evaluation showed the device was returned in original packaging with the batch number of the complaint on the label.The device is fractured at the distal tip.The fractured piece(s) were not returned.Bio debris is present.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the reported event include an inadvertent impact event that may have occurred during device transportation, rough shipping and handling, or at the customer site, as well as the application of an unintended, inappropriate, or excessive force to the device.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|
|
|